CAN/LIN
PC-connectable protocol analyzer / Data logger

LE-270A








●2-way operation of PC-connected protocol analyzer / logger
●2-channel measurement by combination of CAN and LIN
●pass / cut filter and powerful trigger function makes the analysis more efficient
●Robust design to withstand use in harsh field environments

LE-270A is a Communication Data Logger, which records CAN/LIN communication logs in the SD/SDHC card for long hours. Using with a PC, it performs as a CAN/LIN protocol analyzer, and monitors data on the PC at real time (Line Monitor). Using without a PC, it performs as the CAN/LIN data logger, and saves data in the SD/SDHC card. It has two High-speed CAN transceivers, two Low-speed CAN transceivers and two LIN (TJA1020) transceivers. It monitors two channels of CAN or LIN, or one CAN and one LIN transceivers. In addition, four channels of Analog/Digital signals can be recorded.
Power failure protection, water-proofing and low power consumption widens the usage of analyzer such as in the automation factories and other production factories.

  This model is discontinued. The successor is to be released.
Specifications (CAN/LIN Protocol Analyzer / Data Logger LE-270A)
Interface

CAN: Comfort to ISO11898 standard, TJA1050/ISO11519-2, TJA1054. (Switch by the built-in relay)
LIN: Comfort to ISO9141 standard. TJA1021

Connector DSUB 25 pin male connector #4-40UNC
Number of Channels 2 channels of CAN or LIN, or 1 CAN and 1 LIN
Protocol CAN, Device Net, LIN (Rev1.1, 1.2, 1.3, 2.0, 2.1)
Baud Rate CAN: 20Kbps-1Mbps LIN: 400bps-26Kbps (arbitrary)
CAN Monitor Standard/ Expansion format. Support bit timing settings.
LIN Monitor Frame breaking is possible according to the data length of each ID or specified idle time.
Error Check Break:(LIN), Sync:(LIN), Parity:(LIN), Checksum:(CAN/LIN), Framing:(LIN)
Memory

PC: Max 8G byte on the HDD, PC-less: Capacity of the SD card

(Specify the file size as 128K /1M / 2M/ 4M / 8M byte)
Recording Type Ring Buffer (continuous) mode, Fixed Buffer (full stop) mode
Mode Remote mode (with PC); Data Logger mode (PC-less)
Measurement start/stop Control from PC, Start/Stop switch, Auto-Power run, Specify date and time.
Time Stamp “Hr:Min:Sec”, “Min:Sec:x1ms”, 9 digits: “100μs”, “10μs”, “1μs” (selectable)*1
Filter Record specific frames using Bit-mask ID, Bus ID (up to 8), Cut ID (up to 8).
Display on PC Real-time display, Watch data display(display specified data of each ID), Analog waveform display
Trigger Condition Data string up to 8 characters, specified remote frame (CAN), error (Break, Sync, Parity, Checksum, Framing), timer and counter, logic status of external signal.
Action Stop measurement (offset can be set), validates/invalidates trigger condition, control timer/counter, turn on/off the light of user-defined LED, output external signals.
Retrieval function on PC Trigger matched data, Error (Break, Sync, Parity, Checksum, Framing), Data: Specified ID (don’t care available), Data string (Up to 8 characters; don’t care and bit mask available), CAN Remote Data: Specified ID (don’t care available), Specified Time stamp, External signal
External Signal Input Digital/Analog 4 channels
Recording: At the time of receiving signals, or specified sampling cycle (1ms - 10min, 13steps)
Digital VIH 2V (Min.), VIL 0.5V (Max.)
Analog Range: -16V to +16V, Accuracy: ±0.5%FS, A/D conversion: 15Ksps, Resolution: 10bit
Conversion Convert data into Text, CSV
LED 5 two-color LED: Power/Error, Test/Record, CH1/CH2, User-defined U1/U2
Switch One: RUN / STOP
External Trigger Signal 1 Input, 2 Output (equipped in the measurement connector)
SD/SDHC Card  2 – 8G byte*2
USB2.0 Port Mini-B connector. High speed supported.
Power*3

Powered from USB bus power, DC-IN, or 1pin BATTERY terminal
When logger mode: External DC power (DC9-32V), AC adapter (6A-181WP09).

Power consumption: 1.3W (Max. 2.1W for about 10sec after turning on the power). About 0.1W/DC12V while turning off the power.
Run time during power failure Approx. 0.5sec
Ambient Temperature, Humidity In operation: -20 to +60°C, In storage: -20 to +60°C, 20 to 85%RH (No condensation)
Standard CE (Class A)
Dimension, mass 86(W)×130(D)×30(H) mm approx. 240g
Environment of PC OS:Windows® 7/8.1/10  PC: PC/AT compatible
Accessories PC software, Mini-USB cable, CAN/LIN Dsub cable, Power plug cable, 8G byte SDHC card*2

*1:Only “Hr:Min:Sec”, “Min:Sec:x1ms”is available on Analog mode.

*2:Only SDHC card sold by LINEEYE is supported.
*3:AC adapter is sold separately. In the Remote mode (with PC), the analyzer runs by the USB bus power. In the Logger mode (PC-less), you need to have the optional AC adapter (6A-181WP09) or use the proper external power.