OP-8AT

Overview

CAN/high-speed analog measurement expansion set A pod that can be used by attaching to the SB-C2AN to perform high-precision analog measurement and temperature measurement using thermocouples.
* This pod is different from the analog pod (OP-8AH) included with SB-C2AN.


■Specifications
Model OP-8AT
Applicable product SB-C2AN
Channel Number of channels: 8, 24-bit resolution, built-in high-precision ADC, screwless terminal block
Input impedance
voltage resistance
1MΩ, withstand voltage: max±50Vpk, channel-to-channel isolation 350V (AC peak/DC)
Signal isolation voltage Between analog input and analyzer 1500V (AC peak/DC)
350V (AC peak/DC) between each channel of analog input
Input range*1 ±30V, ±10V, ±1V, ±100mV, 0-20mA, temperature K, J, T, E, N, R, S, B type thermocouple compatible
Measurement accuracy※2 ±30V range ±(0.05% rdg + 3mV), ±10V range ±(0.05% rdg + 2mV), ±1V range ±(0.05% rdg + 0.2mV), ±100mV range ±(0.05% rdg + 50µV), 0-20mA current range ±0.05% FS,
Temperature range: See below table
*1: For the 0-20mA range, an external current detection resistor (250Ω or 50Ω, accuracy±0.1% or less) is required on the input terminal block. is.
*2: rdg represents the accuracy for the reading value, and FS represents the accuracy for the full scale of the range. Current accuracy does not include external resistance errors.

OP-8AT temperature measurement specifications
Compatible thermocouple K, J, T, E, N, R, S, B type
Measurement temperature range K type: -200℃ ~ 1370℃
J type: -210℃ ~ 1200℃
T type: -200℃ ~ 400℃
E type: -200℃ ~ 1000℃
N type: -200℃ ~ 1300℃
R type: 0℃ ~ 1760℃
S type: 0℃ ~ 1760℃
B type: 400℃ ~ 1800℃
Cold junction compensation Internal compensation, external compensation switchable
Disconnection detection function On/off switchable (applied current: approx. 180nA)
Measurement accuracy*1 K type -50℃ ~ 1370℃ : ±(0.05% rdg + 1.0℃)
-200℃ ~ -50℃ : ±(0.05% rdg + 2.0℃)
J type -50℃ ~ 1200℃ : ±(0.05% rdg + 0.8℃)
-210℃ ~ -50℃ : ±(0.05% rdg + 1.6℃)
T type -50℃ ~ 400℃ : ±(0.05% rdg + 1.0℃)
-200℃ ~ -50℃ : ±(0.05% rdg + 2.0℃)
E type -50℃ ~ 1000℃ : ±(0.05% rdg + 0.6℃)
-200℃ ~ -50℃ : ±(0.05% rdg + 1.2℃)
N type -50℃ ~ 1300℃ : ±(0.05% rdg + 1.5℃)
-200℃ ~ -50℃ : ±(0.05% rdg + 3.0℃)
R type
S type
400℃ ~ 1760℃ : ±(0.05% rdg + 3.5℃)
0℃~400℃: ±(0.05% rdg + 6.0℃)
B type 800℃ ~ 1800℃ : ±(0.05% rdg + 4.0℃)
400℃ ~ 800℃ : ±(0.05% rdg + 7.5℃)
Cold junction compensation accuracy*2 ±1.0℃
*1:  Accuracy at ambient temperature of 18 to 28°C and after 20 minutes after power on, does not include thermocouple error.
If the above ambient temperature range is exceeded, add 1/20 of each error value for every 1°C.
*2:  In environments where a part of the measurement pod is locally heated or cooled, cold junction compensation accuracy cannot be guaranteed.